11.4 ELECTROMAGNETIC IMMUNITY
The device is intended for use in the following electromagnetic environment. You must ensure
compliance with this guideline.
Table 23 - Electromagnetic immunity
CAUTION
Interference may occur in the vicinity of equipment marked with the following symbol:
GUIDANCE AND MANUFACTURER DECLARATION
ELECTROMAGNETIC IMMUNITY
The EMBODY
is intended for use in the electromagnetic environment specified below.
The operator of the EMBODY
must ensure that it is used in such an environment.
Immunity test CEI 60601 test level
Compliance
level
Electromagnetic
environment - Guidance
Electrostatic discharge
(ESD) acc. To
IEC 61000-4-2
- ±8 kV contact
discharge
- ± 2 kV ,± 4 kV,± 6
kV,± 8 kV,± 15 kV air
discharge
- ±8 kV contact
discharge
- ± 2 kV,± 4 kV,±
6 kV,± 8 kV,± 15
kV air
discharge
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic materials, the
relative humidity should be
at least 30%.
Electrical fast
transients/ burst acc.
To
IEC 61000-4-4
- ±2 kV/100Hz for
power supply lines
±2 kV/100Hz for
power supply
lines
The quality of the supply
voltage should correspond
with one characteristic for a
typical commercial or
hospital environment.
Surge acc. To
IEC 61000-4-5
- ±0.5, ±1 kV
differential mode
±0.5, ±1 kV
differential mode
The quality of the supply
voltage should correspond
with one characteristic for a
typical commercial or
hospital environment.
Voltage dips, short-
term interruptions and
voltage variations on
power supply input
lines acc. to 61000-4-
11
<5% U
T
during 0,5
period
40% U
T
during 5
periods
70% U
T
during 25
periods
<5% U
T
during 5s
Compliant to
the specified
levels
Compliant
The quality of the supply
voltage should correspond
to one characteristic for a
typical commercial or
hospital environment. If the
user of the EMBODY
requires a continuous
function of the appliance
also during interruptions of
the power supply, it is
recommended to power
EMBODY out of an
uninterruptible power
supply or a battery.
Power frequency
(50/60Hz) magnetic
fields acc. to IEC
61000-4-8
30 A/m 30 A/m
Power frequency magnetic
fields should be at levels
characteristic for
commercial or hospital
environments.
NOTE: U
T
is the voltage of the alternative supply voltage before the application of the test level