TROUBLESHOOTING
If you have trouble in using the unit, please check the following points first.
Note: Relax for at least 10 minutes and then take another measurement.
SPECIFICATIONS
*
Specifications may be changed without notice in the event of improvement
being made.
Measuring Method:
Indication:
Measuring Range:
Accuracy:
Inflation
Deflation:
Memory:
Power Source:
Battery Life:
Working temp. & Humidity:
Storage Temp. & Humidity:
Outside Dimensions:
Weight:
Accessories:
Operation Voltage 6 V DC
Oscillometric System
Digital LCD Display
Pressure: 0~300mmHg, Pulse: 40~200 Beat/min.
Pressure ±3mmHg; Pulse ±5%
Electric Pump Inflation
Pressure Release Valve
128 Memories
1.5V "AA" x 4 Alkaline Batteries
Approx. 250 Times (Once a Day, Room Temperature 22
O
C)
40~104
O
F (5~40
O
C), 30~85% RH
4~140
O
F (-20~60
O
C), 10~95% RH
112mm(W) x 140mm(D) x 70mm(H)
Approx. 534g (Apparatus and Battery)
Cuff (fit arm size 22~32cm) , Instruction Manual
ERROR DISPLAY
Nothing is displayed
when you push the
POWER button or
battery icon flash
E1: Can't normally
increase pressure
E2, E7,E8,E9:
Measurement failure
E3. Inflate pressure too high
E4: Have shaking while
measurement
E5: Automatically pressure
release too fast
E6: Automatically pressure
release too slow
Battery icon on
The systolic pressure
value or diastolic
pressure value too high
The systolic pressure
value or diastolic
pressure value too low
POSSIBLE CAUSE
No battery installation
Battery worn out
The polarities of batteries
placed wrongly
Check your cuff if any air
leakage
Arm cuff not properly adjust
or placement, hand or body
shaking
Hand or body shaking while
measurement
Cuff too tight
Cuff too loose
Battery low power
1. The cuff was held lower
than your heart
2. The cuff was not
attached properly
3. You moved your body or
spoke during measurement
1. The cuff was held higher
than your heart
2. You moved your body or
spoke during measurement.
HOW TO CORRECT
Insert batteries
Replace new batteries
Insert battery in the correct
polarities
Replace cuff with new one
Renew adjust cuff position and
tightness, and keep hand still
Re-measurement or send back
dealer for re-calibrate pressure
Keeping static and correct gesture
to measure again
Readjust cuff and measure again
Replace battery, and measure again
Keeping correct position and
gesture to measure again