Operation Manual

If you have trouble while using the unit, please check the following points first.
TROUBLESHOOTING
ERROR DISPLAY POSSIBLE CAUSE HOW TO CORRECT
Nothing is displayed
when you push the
POWER button or
battery icon flash
The polarities of batteries
placed wrongly
Battery worn out
No battery installation Insert batteries
Replace new batteries
Insert battery in the
correct polarities
E1: Can't normally
increase pressure
Check your cuff if any air
leakage
Replace wrist cuff with
new one
E0, E2, E7, E8, E9
Measurement failure
Cuff not properly adjust or
placement, hand or body
shaking.
Renew adjust cuff position
and tightness, and keep
hand still
E3. Inflate pressure
too high
Re-measurement or send back
dealer for re-calibrate pressure
E4: Have shaking
while measurement
Hand or body shaking while
measurement
Keeping static and correct
gesture to measure again
E5: Automatically
pressure release
too fast
Cuff too tight
E6: Automatically
pressure release
too slow
Cuff too loose
Readjust cuff and measure
again
" " Battery icon on
Battery low power
Replace battery, and
measure again
The systolic pressure
value or diastolic
pressure value too
high
1. The hand with the cuff was
held lower than your heart
2. The cuff was not attached
properly
3. You moved your body or
spoke during measurement
Keeping correct position and
gesture to measure again
1. The hand with the cuff was
held higher than your heart
2. You moved your body or
spoke during measurement.
The systolic pressure
value or diastolic
pressure value too
low
Note: Relax for at least 10 minutes and then take another measurement.
-17-
611-200600-000