Operation Manual

Measuring Method:
Indication:
Measuring Range:
Accuracy:
Inflation
Deflation:
Memory:
Power Source:
Battery Life:
Working Temp. & Humidity:
Storage Temp. & Humidity:
Outside Dimensions:
Weight:
Accessories:
Oscillometric System
Digital LCD Display
Pressure: 0~300 mmHg;
Pressure: ±3 mmHg; Pulse ±5%
Electric Pump Inflation
Pressure Release Valve
128 Memories
1.5V "AAA" x 2 Alkaline Battery
Approx. 250 Times (Once a Day, Room
Temperature 22
o
C)
40~104
o
F (5~40
o
C), 30~85% RH
4~140
o
F (-20~60
o
C), 10~95% RH
67mm(W) x 76mm(D) x 32mm(H)
Approx. 128g (Apparatus and Battery)
Wrist Cuff (Fit size 13.5-19.5cm),
Operation Voltage 3V DC
-21-
SPECIFICATIONS
Pulse: 40~200 Beat/min.
*Specitications may be changed Without notice in the event of
lmprovement being made.
TROUBLESHOOTING
ERROR DISPLAY
Nothing is displayed
when you push the
POWER button or
batter yicon flash
E1: Can't normally
increase pressure
E2, E7, E8, E9
Measurement failure
E3. Inflate pressure
too high
E4: Have shaking while
measurement
E5: Automatically pressure
release too fast
E6: Automatically pressure
release too slow
" " Battery icon on
The systolic pressure
value or diastolic
pressure value too high
The systolic pressure
value or diastolic
pressure value too low
POSSIBLE CAUSE
No battery installation
Battery worn out
The polarities of batteries
placed wrongly
Check your cuff if any air
leakage
Wrist cuff not properly
adjust or placement,
hand or body shaking.
Hand or body shaking while
measurement
Cuff too tight
Cuff too loose
Battery low power
1. The hand with the cuff was
held lower than your heart
attached properly
3. You moved your body or
spoke during measurement
HOW TO CORRECT
Insert batteries
Replace new batteries
Insert battery in the correct
polarities
Replace wrist cuff with new one
Renew adjust wrist cuff position
and tightness, and keep hand still
Re-measurement or send back
dealer for re-calibrate pressure
Keeping static and correct gesture
to measure again
Readjust cuff and measure again
Replace battery, and measure again
Keeping correct position and
gesture to measure again
-22-
If you have trouble while using the unit, please check the
following points first.
Instruction Manual
2. The cuff was not
1. The hand with the cuff was
held higher than your heart
2. You moved your body or
spoke during measurement.
Note: 1) Relax for at least 10 minutes and then take another
measurement.