User's Manual

Smart Storage Agitator for PC4200 User Manual
Version: 2.2
Validation date of the manual: 31/10/2018 10 of 33
The operating principle of the RFID system is based on a transponder (RFID tags) and an
interrogator (coupler). The latter is an active radiofrequency transmitter device that will
activate the RFID tags located in the slot by providing them with the energy they need to
operate. In addition to the energy, the interrogator sends particular commands to which the
RFID tag responds. A simple command may be to return the matching donation number to a
unique identifier.
2.3.4. Temperature sensor
The temperature sensor integrated in the SST-A for PC4200 is waterproof. It allows to
measure the temperature of an area.
Fig. Temperature sensor
Only the temperature of the climatic chamber (incubator) is the reference. SST-A for PC4200
has no claimed performance regarding temperature.
The SST-A for PC4200 gives the temperature only as an indication. This function does not
provide the safety associated with maintaining the storage temperature of platelet
concentrates.
2.3.5. Agitation sensor
The Agitation sensor (accelerometer) integrated in the SST-A for PC4200 set allows to record
the shaking and stopping phases of the agitator.
2.3.6. Shielding
Fig. Upper shielding