User's Manual

Smart Storage Agitator for PC4200 User Manual
Version: 2.2
Validation date of the manual: 31/10/2018 21 of 33
Certain types of mobile telecommunication devices such as mobile phones are likely to
interfere with the SST-A. The separation distances recommended in this chapter must
therefore be complied with.
The SST-A should not be used in the vicinity or placed on top of another device. If this cannot
be avoided, its effective operation must be checked under operational conditions before
use. Using accessories other than those specified or sold by Biolog-id as replacement parts
can increase emissions or reduce the immunity of the SST-A.
The SST-A uses the 13.56 MHz frequency. The frequency band is 13.553 - 13.567 MHz in
accordance with the ISO 15693 standard. The modulation type is ASK and the RF mode is
TX/RX.
The maximum power output of the PRD_7150400A is -6.40dBµA/m at 3m.
All information featured below comes from normative requirements which apply to the
manufacturers of medical electrical devices, under standard IEC60601-1-2 Ed4.
Length of cables:
Cables and accessories
Maximum length
Type of test
In accordance with:
Power cable
CAN cable
Ethernet cable
< 3m
< 3m
> 3m
RF emission
CISPR 11, Class B
Harmonic current emissions
IEC 61000-3-2
Voltage fluctuation and flicker
IEC 61000-3-3
Electrostatic discharge immunity
IEC 61000-4-2
Radiated immunity
Electromagnetic fields
IEC 61000-4-3
Immunity to fast transient bursts
IEC 61000-4-4
Surge immunity
IEC 61000-4-5
Conducted immunity Conducted
radio frequency interference
IEC 61000-4-6
Radiated immunity - Magnetic
fields
IEC 61000-4-8
Voltage dips, short interruptions
and voltage variations immunity
IEC 61000-4-11