User's Manual

Smart Storage Agitator for PC4200 User Manual
Version: 2.2
Validation date of the manual: 31/10/2018 23 of 33
Magnetic and electromagnetic immunity
The SST-A is intended to be used in the magnetic and electromagnetic environment
described in the table below. The user and the installer must guarantee the compliance of
the electromagnetic environment.
Immunity test
Test level according to
IEC60601
Compliance level
Electromagnetic
environment/comments
Electrostatic discharge
(ESD)
(IEC61000-4-2)
± 8 kV contact discharge
± 2 kV; ± 4 kV; ± 8 kV; ± 15
kV air discharge
± 8 kV contact discharge
± 15 kV air discharge
Professional healthcare
facility environment.
Fast transient bursts
(IEC61000-4-4)
± 2 kV for power supply
lines
± 2 kV for power supply
lines
± 1 kV for signal ports
Professional healthcare
facility environment.
Surges
(IEC61000-4-5)
± 1 kV in Differential mode
± 2 kV in common mode
± 1 kV in Differential mode
± 2 kV in common mode
Professional healthcare
facility environment.
Power-frequency magnetic
field
(IEC61000-4-8)
30 A/m
30 A/m
Professional healthcare
facility environment.
Voltage dips, short
interruptions and voltage
variations
(IEC61000-4-11)
0% UT for 0.5 cycles
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
0% UT for 1 cycle
70% UT
for 25 cycles at 50 Hz
For 30 cycles at 60 Hz
Single-phase: at 0°
0% UT for 0.5 cycles
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
0% UT for 1 cycle
70% UT
for 25 cycles at 50 Hz
For 30 cycles at 60 Hz
Single-phase: at 0°
Professional healthcare
facility environment.
Voltage interruptions
(IEC61000-4-11)
0 % UT;
for 250 cycles at 50 Hz
for 300 cycles at 60 Hz
0 % UT;
for 250 cycles at 50 Hz
for 300 cycles at 60 Hz
Professional healthcare
facility environment.