Electromagnetic immunity, radio frequencies:
The SST-A is intended to be used in the magnetic and electromagnetic environment
described in the table below. The user and the installer must guarantee the compliance of
the electromagnetic environment.
Immunity test
Test level
Compliance level
Electromagnetic
environment/comments
WARNING: Portable RF communication devices (including peripherals such as antenna cables and
external antennas) should not be used within 30 cm (12 inches) of any part of the PRD_7150300A,
PRD_7150400A, PRD_7150500A and PRD_7150600A, including cables specified by the
manufacturer. Otherwise the performance of these devices could be adversely affected.
Radiated RF
electromagnetic fields
(IEC61000-4-3)
3 V/m
80 MHz to 2.7 GHz
80% MA at 1 kHz
3 V/m
80 MHz to 2.7 GHz
80% MA at 1 kHz
Professional healthcare
facility environment.
Proximity fields emitted by
wireless RF communication
devices (IEC 61000-4-3
interim
method)
9 V/m
710 MHz, 745 MHz,
780 MHZ, 5240 MHz,
5550 MHz, 5785 MHz
27 V/m
385 MHz
28 V/m
450 MHz, 810 MHz,
870 MHz, 930 MHz,
1720 MHz, 1845 MHz,
1970 MHz, 2450 MHz
9 V/m
710 MHz, 745 MHz,
780 MHZ, 5240 MHz,
5550 MHz, 5785 MHz
27 V/m
385 MHz
28 V/m
450 MHz, 810 MHz,
870 MHz, 930 MHz,
1720 MHz, 1845 MHz,
1970 MHz, 2450 MHz
Professional healthcare
facility environment.
Conducted disturbances,
induced by RF
fields (IEC610004-6)
3 V
150 KHz to 80 MHz
6 V within ISM band and
3 V
150 KHz to 80 MHz
6 V within ISM band and
Professional healthcare
facility environment.