4.2 RF radiation hazards.
The SST-F electronic system antennas each emit a frequency of 13.56 MHz
with a maximum power output of -2.38dBµA/m at 3m (less than the
42dBµA/m limit threshold).
MEDICAL ELECTRICAL DEVICES require special EMC precautions. The SST-
F must be installed and commissioned in accordance with the EMC data
supplied in the ACCOMPANYING DOCUMENTS.
Portable or mobile RF communication devices can affect MEDICAL
ELECTRICAL DEVICES
Using ACCESSORIES, transducers or cables other than those specified can
increase EMISSIONS or reduce the IMMUNITY of the DEVICE or EM
SYSTEM. This does not include transducers and cables sold by the
MANUFACTURER of the DEVICE or EM SYSTEM and used as spare parts to
replace internal components.
The DEVICE or EM SYSTEM must not be used beside other devices or
stacked on top of them.
The DEVICE or EM SYSTEMS can be affected by interference caused by
other devices even if they comply with CISPR EMISSION requirements.
4.3 Electromagnetic compatibility:
While the SST-F complies with current electromagnetic compatibility standards, users must
ensure that any electromagnetic interference does not create additional risk, such as RF
transmitters or other electronic devices.
In this chapter you will find the information you need to ensure the best possible installation
and use of the SST-F in terms of electromagnetic compatibility.
The different cords of the SST-F must be placed far apart.
Certain types of mobile telecommunication devices such as mobile phones are likely to
interfere with the SST-F. The separation distances recommended in this chapter must
therefore be complied with.
The SST-F should not be used in the vicinity or placed on top of another device. If this cannot
be avoided, its effective operation must be checked under operational conditions before use.
Using accessories other than those specified or sold by Biolog-id as replacement parts can
increase emissions or reduce the immunity of the SST-F.
The SST-F uses the 13.56 MHz frequency. The frequency band is 13.553 - 13.567 MHz in
accordance with the ISO 15693 standard. The modulation type is ASK and the RF mode is
TX/RX.
The maximum power output of the PRD-7140100A and PRD_7140300A is -2.38dBµA/m at 3m.
The tables below relate to the SST-F (PRD-7140100A).