User's Manual

Smart Storage Freezer user manual
Version
4.0
Manual validation date: 18/07/2019
21 of 33
Directives and manufacturer declaration - electromagnetic immunity
The PRD-7140100A is designed to be used in the electromagnetic environment specified below. PRD-7140100A
customers or users must ensure that their devices are used in such an environment.
Immunity test
Test level
CEI 60601
Level of
compliance
Electromagnetic environment - directives
Voltage dips, short
outages and voltage
fluctuations on
power supply input
lines
CEI 61000-4-11
<5% UV
(>95% UV dip) for 0.5
cycles
40 % UV
(60 % UV dip) for 5
cycles
70 % UV
(30% UV dip) for 25
cycles
<5% UV
(>95% UV dip) for 5
seconds
<5% UV
(>95% UV dip) for
0.5 cycles
40 % UV
(60 % UV dip) for 5
cycles
70 % UV
(30% UV dip) for 25
cycles
<5% UV
(>95% UV dip) for 5
seconds
The quality of the power supply network must be
equivalent to that of a standard retail or hospital
environment. If a PRD-7140100A user requires
continuous operation during electrical power
supply network outages, we recommend
connecting the PRD-7140100A to an
uninterruptible power supply or battery.
Power-frequency
magnetic field (50/60
Hz)
CEI 61000-4-8
3 A/m
3 A/m
Power-frequency magnetic fields must remain at
levels characteristic of a representative location
in a typical retail or hospital environment.
NOTE UV is the voltage of the AC network before application of the test level.
Conducted RF
emissions
CEI 61000-4-6
Radiated RF
emissions
CEI 61000-4-3
3 Veff
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 V
3 V/m
Portable or mobile RF communication devices
must not be used any closer than the
recommended separation distance from any
part of the PRD-7140100A, including the cables,
as calculated using the applicable equation for
the emitter frequency.
Recommended separation distance
d = 1.17 √ P
d = 1.17 √ P 80 MHz to 800 MHz
d = 2.34 √ P 800 MHz to 2.5 GHz
where P is the maximum characteristic power
output of the emitter in Watts (W) as provided by
the manufacturer of the emitter, and d is the
recommended separation distance in metres
(m).
The strength of fixed RF emitter fields,
determined by carrying out an electromagnetic
survey on site , must be below the compliance
level in each of the range of frequencies.
Interference can be caused by proximity to
devices marked with the following symbols: