Immunity test
Test level according to
IEC60601
Compliance level
Electromagnetic
environment/comments
Electrostatic discharge
(ESD)
(IEC61000-4-2)
± 8 kV contact discharge
± 2 kV; ± 4 kV; ± 8 kV; ± 15
kV air discharge
± 8 kV contact discharge
± 15 kV air discharge
Professional healthcare
facility environment.
Fast transient bursts
(IEC61000-4-4)
± 2 kV for power supply
lines
± 2 kV for power supply
lines
± 1 kV for signal ports
Professional healthcare
facility environment.
Surges
(IEC61000-4-5)
± 1 kV in Differential mode
± 2 kV in common mode
± 1 kV in Differential mode
± 2 kV in common mode
Professional healthcare
facility environment.
Power-frequency magnetic
field
(IEC61000-4-8)
30 A/m
30 A/m
Professional healthcare
facility environment.
Voltage dips, short
interruptions and voltage
variations
(IEC61000-4-11)
0% UT for 0.5 cycles
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
0% UT for 1 cycle
70% UT
for 25 cycles at 50 Hz
For 30 cycles at 60 Hz
Single-phase: at 0°
0% UT for 0.5 cycles
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
0% UT for 1 cycle
70% UT
for 25 cycles at 50 Hz
For 30 cycles at 60 Hz
Single-phase: at 0°
Professional healthcare
facility environment.
Voltage interruptions
(IEC61000-4-11)
0 % UT;
for 250 cycles at 50 Hz
for 300 cycles at 60 Hz
0 % UT;
for 250 cycles at 50 Hz
for 300 cycles at 60 Hz
Professional healthcare
facility environment.
Electromagnetic immunity, radio frequencies:
The SST-F is intended to be used in the magnetic and electromagnetic environment
described in the table below. The user and the installer must guarantee the compliance of the
electromagnetic environment.