User's Manual

Smart Storage Freezer user manual
Version
4.0
Manual validation date: 18/07/2019
27 of 33
Immunity test
Test level
Compliance level
Electromagnetic
environment/comments
WARNING: Portable RF communication devices (including peripherals such as antenna cables and
external antennas) should not be used within 30 cm (12 inches) of any part of the PRD_7140300A,
including cables specified by the manufacturer. Otherwise the performance of these devices could
be adversely affected.
Radiated RF
electromagnetic fields
(IEC61000-4-3)
3 V/m
80 MHz to 2.7 GHz
80% MA at 1 kHz
3 V/m
80 MHz to 2.7 GHz
80% MA at 1 kHz
Professional healthcare
facility environment.
Proximity fields emitted by
wireless RF communication
devices (IEC 61000-4-3
interim
method)
9 V/m
710 MHz, 745 MHz,
780 MHZ, 5240 MHz,
5550 MHz, 5785 MHz
27 V/m
385 MHz
28 V/m
450 MHz, 810 MHz,
870 MHz, 930 MHz,
1720 MHz, 1845 MHz,
1970 MHz, 2450 MHz
9 V/m
710 MHz, 745 MHz,
780 MHZ, 5240 MHz,
5550 MHz, 5785 MHz
27 V/m
385 MHz
28 V/m
450 MHz, 810 MHz,
870 MHz, 930 MHz,
1720 MHz, 1845 MHz,
1970 MHz, 2450 MHz
Professional healthcare
facility environment.
Conducted disturbances,
induced by RF
fields (IEC610004-6)
3 V
150 KHz to 80 MHz
6 V within ISM band and
band ranging from
0.15 MHZ to 80 MHZ,
80% MA at 1 KHz
3 V
150 KHz to 80 MHz
6 V within ISM band and
band ranging from
0.15 MHZ to 80 MHZ,
80% MA at 1 KHz
Professional healthcare
facility environment.