User's Manual

Smart Storage High Density User Manual
Version
0.6
Manual validation date: 04/09/2020
19 of 31
4.3 Electromagnetic compatibility:
While the SST-HD complies with current electromagnetic compatibility standards, users
must ensure that any electromagnetic interference does not create additional risk, such as
RF transmitters or other electronic devices.
In this chapter you will find the information you need to ensure the best possible installation
and commissioning of the SST-HD in terms of electromagnetic compatibility.
The different cords of the SST-HD must be placed far apart.
Certain types of mobile telecommunication devices such as mobile phones are likely to
interfere with the SST-HD. The separation distances recommended in this chapter must
therefore be complied with.
The SST-HD should not be used in the vicinity or placed on top of another device. If this
cannot be avoided, its effective operation must be checked under operational conditions
before use. Using accessories other than those specified or sold by Biolog-id as replacement
parts can increase emissions or reduce the immunity of the SST-HD.
The SST-HD uses the 13.56 MHz frequency. The frequency band is 13.553 - 13.567 MHz in
accordance with the ISO 15693 standard. The modulation type is ASK and the RF mode is
TX/RX.
The maximum power output of the PRD_7170100A and PRD_7170200A is 6.9dBµA/m at 3m.
The tables below relate to the SST-HD range (PRD_7170100A et PRD_7170200A).
All information featured below comes from normative requirements which apply to the
manufacturers of medical electrical devices, under standard IEC60601-1-2 Ed4.