User's Manual

Smart Storage High Density User Manual
Version
0.6
Manual validation date: 04/09/2020
20 of 31
Length of cables:
Cables and accessories
Maximum
length
Type of test
In accordance with
Power cable
CAN cable
Ethernet cable
< 3m
< 3m
> 3m
RF emission
CISPR 11, Class B
Harmonic current emissions
IEC 61000-3-2
Voltage fluctuation and flicker
IEC 61000-3-3
Electrostatic discharge immunity
IEC 61000-4-2
Radiated immunity Electromagnetic
fields
IEC 61000-4-3
Immunity to fast transient bursts
IEC 61000-4-4
Surge immunity
IEC 61000-4-5
Conducted immunity Conducted radio
frequency interference
IEC 61000-4-6
Radiated immunity - Magnetic fields
IEC 61000-4-8
Voltage dips, short interruptions and
voltage variations immunity
IEC 61000-4-11
Recommended separation distances:
The SST-HD is intended to be used in an electromagnetic environment in which radiated RF
disturbances are controlled.
The user or installer of the medical device can help prevent any electromagnetic interference
by maintaining a minimum distance, as a function of the maximum power output of the radio
frequency transmission equipment. Portable RF communication devices (including
peripherals such as antenna cables and external antennas) should not be used within 30 cm
(12 inches) of any part of the SST-HD, including cables specified by the manufacturer.
Otherwise the performance of these devices could be adversely affected.
Electromagnetic emissions
The SST-HD is intended to be used in the electromagnetic environment described in the table
below. The user and installer must therefore ensure that the SST-HD is used in the
environment described below.