Emission test
Compliance
Electromagnetic environment - comments
Radiated electromagnetic
disturbance
(Radiated emissions)
(CISPR 11)
Group 1
The PRD_7170100A et PRD_7170200A medical
devices uses RF energy for internal operation.
Power terminal disturbance voltage
(Conducted emissions)
(CISPR 11)
Class B
NA
Harmonic current emissions
(IEC61000-3-2)
Compliant
Voltage variations, voltage
fluctuations and flicker
(IEC61000-3-3)
Compliant
Magnetic and electromagnetic immunity
The SST-HD is intended to be used in the magnetic and electromagnetic environment
described in the table below. The user and the installer must guarantee the compliance of the
electromagnetic environment.
Immunity test
Test level according to
IEC60601
Compliance level
Electromagnetic
environment/comments
Electrostatic discharge (ESD)
(IEC61000-4-2)
± 8 kV contact discharge
± 2 kV; ± 4 kV; ± 8 kV; ± 15
kV air discharge
± 8 kV contact discharge
± 15 kV air discharge
Professional healthcare
facility environment.
Fast transient bursts
(IEC61000-4-4)
± 2 kV for power supply
lines
± 2 kV for power supply
lines
± 1 kV for signal ports
Professional healthcare
facility environment.
Surges
(IEC61000-4-5)
± 1 kV in Differential mode
± 2 kV in common mode
± 1 kV in Differential mode
± 2 kV in common mode
Professional healthcare
facility environment.
Power-frequency magnetic
field
(IEC61000-4-8)
30 A/m
30 A/m
Professional healthcare
facility environment.
Voltage dips, short
interruptions and voltage
variations
(IEC61000-4-11)
0% UT for 0.5 cycles
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
0% UT for 0.5 cycles
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 31
Professional healthcare
facility environment.