User's Manual
Smart Storage High Density User Manual
Version
0.6
Manual validation date: 04/09/2020
22 of 31
0% UT for 1 cycle
70% UT
for 25 cycles at 50 Hz
For 30 cycles at 60 Hz
Single-phase: at 0°
0% UT for 1 cycle
70% UT
for 25 cycles at 50 Hz
For 30 cycles at 60 Hz
Single-phase: at 0°
Voltage interruptions
(IEC61000-4-11)
0% UT;
for 250 cycles at 50 Hz
for 300 cycles at 60 Hz
0% UT;
for 250 cycles at 50 Hz
for 300 cycles at 60 Hz
Professional healthcare
facility environment.
Electromagnetic immunity, radio frequencies:
The SST-HD is intended to be used in the magnetic and electromagnetic environment
described in the table below. The user and the installer must guarantee the compliance of the
electromagnetic environment.
Immunity test
Test level
Compliance level
Electromagnetic
environment/comments
WARNING: Portable RF communication devices (including peripherals such as antenna cables and
external antennas) should not be used within 30 cm (12 inches) of any part of the PRD_7170100A
and PRD_7170200A, including cables specified by the manufacturer. Otherwise the performance of
these devices could be adversely affected.
Radiated RF
electromagnetic fields
(IEC61000-4-3)
3 V/m
80 MHz to 2.7 GHz
80% MA at 1 kHz
3 V/m
80 MHz to 2.7 GHz
80% MA at 1 kHz
Professional healthcare
facility environment.
Proximity fields emitted by
wireless RF communication
devices (IEC 61000-4-3
interim method)
9 V/m
710 MHz, 745 MHz,
780 MHz, 5240 MHz,
5550 MHz, 5785 MHz
27 V/m
385 MHz
28 V/m
450 MHz, 810 MHz,
9 V/m
710 MHz, 745 MHz,
780 MHz, 5240 MHz,
5550 MHz, 5785 MHz
27 V/m
385 MHz
28 V/m
450 MHz, 810 MHz,
Professional healthcare
facility environment.