User's Manual

Smart Storage High Density User Manual
Version
0.6
Manual validation date: 04/09/2020
28 of 31
Red and green indicators
on the front of the
processor
Inform the member of staff in charge of
maintenance so that corrective maintenance
can be carried out.
* In the event of a failure, RFID traceability may be interrupted. This interruption is referred to in the
product’s event log.
When a red LED appears, try to detect the cause of the fault and eliminate it as quickly as
possible.
Red indicator on satellite
Possible causes
Actions
Data cannot be written
Move the bag to a different space and try again
The tray has been open for more than 4
minutes
Close the tray
Communication with processor lost
Inform the member of staff in charge of
maintenance so that corrective maintenance
can be carried out.
7 Warranty
Failure to observe any of the recommendations will void the warranty.
8 Transport
When receiving the SST-HD, verify that it has not been damaged during transport.
If you notice any transport-related anomalies, immediately contact your carrier or retailer
and show them the delivery note or purchase order.
The required transport conditions are determined by Biolog-id. They must be respected to
preserve the physical integrity of the device.
9 Manufacturer liability
The manufacturer shall not be held liable in the following cases:
Failure to observe the manufacturer’s installation recommendations.
Table 2 Symptoms or faults: Red indicator on the satellite