User's Guide

13
ment or after charging and can reach up to 41°C. To prevent overheat-
ing and prolong its life let the device cool down to ambient tempera-
ture before use.
CAUTION: Do not use the OrthoPulse® device for more than
two 5-minute sessions subsequently. It may cause the device to
overheat.
CAUTION: Ensure the device is not warm after it is charged
before starting treatment.
Transport and Storage Conditions:
Minimum ambient temperature: -25°C
Maximum ambient temperature: 70°C
Maximum humidity: 90% non-condensing
Storage pressure range 700 to 1060 hPa
WARNING: The device shall not be used in conditions beyond
the listed above only! The patient shall check the device for any
visual defects prior to performing the treatment.
5.3 EMC Compliance Statement
This device has been tested and found to comply with the limits for a
Class B digital device, pursuant to part 15 of the FCC rules. These limits
are designed to provide reasonable protection against harmful interfer-
ence in a residential setting. This device generates, uses and can radi-
ate radio- frequency energy and, if not installed and used in accordance
with the instructions, may cause harmful interference to radio com-
munications. However, there is no guarantee that interference will
not occur in a particular installation. If this device does cause harmful
interference to radio or television reception, which can be determined
by turning the device off and on, the user is encouraged to try one or
more of the following measures:
Reorient or relocate the device or the receiver
Increase the distance between the device and the receiver
Connect the device to an outlet on a circuit different from that to
which the receiver is connected
Consult the manufacturer or an experienced broadcast engineer/
technician for help
Be aware that portable and mobile radio-frequency communications
equipment (for example, mobile phones, iPads) may affect the opera-
tion of this device; take appropriate precautions during operation.
Accessories
To maintain electromagnetic compatibility (EMC) within limits, the
device must be used with the cables and accessories specified by Biolux.
The use of accessories or cables other than those specified or supplied
may result in increased emissions or decreased immunity of the device.
Radio-Frequency Transmitter
OrthoPulse® contains a Bluetooth LE transmitter module that oper-
ates at 2.4 GHz. This module is active only when the device is placed in
the charging case and the Ready for Bluetooth indicator is on.
This equipment has been tested and found to comply with the
limits for a Class B digital device, pursuant to part 15 of the FCC Rules.
These limits are designed to provide reasonable protection against
harmful interference in a residential installation.