Test or Measurement Standards Test Method Description Results
Radiated Emissions EN 60601-1-2:2015
Ed. 4
CISPR 11
EN 60601-1-11
EN 60601-2-57
EN 301 489-1
V2.1.1
ICES-003 Issue 6
CFR Title 47 FCC Part 15
ICES-003 Issu.6
Class B Limits
The radiated emissions
are measured in the
30-1000MHz range or up
to 5x the highest EUT fre-
quency whichever is higher
1
Complies
Conducted Emissions The Conducted Emissions
are measured on the phase
and Neutral Power lines in
the 0.15 - 30.0 MHz range.
1
Highest frequency generated by the device is 2.4GHz
Emission Test Compliance Comments
RF Emissions CISPR 11 Group 1 This device uses RF energy only for its internal
functions.
RF Emissions CISPR 11 Class B
This device is predominantly intended for use
in a HOME HEALTHCARE ENVIROUNMNET
and to be connected to the PUBLIC
MAINS NETWORK
Harmonic Emissions EN 61000-3-2
Class A
Voltage Fluctuations/
Flicker Emissions
EN 61000-3-3
Class A