Immunity Test Standard/Test Method Test Levels Compliance
Electrostatic Discharge IEC 61000-4-2 Air Discharge: ± 2, 4, 8, 15 kV
Contact Discharge: ± 8 kV
Complies
Radiated RF IEC 61000-4-3 10V/m, 80% AM @
1kHz, 30MHz to 2.5GHz,
Vertical and Horizontal
Polarizations
Complies
Immunity to Proximity Fields from RF Wireless
Communications Equipment
IEC 61000-4-3 9 V/m to 28 V/m @ 15
Frequencies 380 - 5800 MHz
Complies
Electrical Fast Transient/ Burst IEC 61000-4-4 AC Power Lines: ± 2 kV @ 100 kHz
Signal Lines: ± 1 kV @ 100 kHz
Complies
Surge IEC 61000-4-5 ±0.5, 1 kV line to line, 0°, 90°,180°, 270°
±0.5. 1, 2 kV line to earth, 0°, 90°,180°, 270°
Complies
Conducted RF IEC 61000-4-6 3Vrms, 0.15-80MHz, 80%
AM @ 1 kHz
6Vrms in ISM & Amateur
radio bands, 0.15-80MHz,
80% AM @ 1 kHz
Complies
Power Frequency Magnetic
Field
IEC 61000-4-8 30 A/m Complies
Voltage Dips IEC 61000-4-11 0 % UT; 0,5 cycle at 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
0 % UT; 1 cycle
70 % UT; 25/30 cycles
Complies
Voltage Interruptions IEC 61000-4-11 0 % UT; 250/300 cycles Complies
NOTE: UT is the AC mains voltage prior to application of the test level.
Electromagnetic Immunity*