User's Guide

7
6 h
5 min
5 min
You may pause the treatment
for up to 20 seconds by simply
removing the device from your
mouth. If you pause for more than
20 seconds, the treatment will
abort and you will have to restart
your OrthoPulse® treatment.
Applied Part
To use your OrthoPulse®,
complete the six steps below:
1. Remove the OrthoPulse® from the charging case,
this will wake the device from sleep mode. The
status light will display green upon waking when
the battery has sufficient charge to complete a
treatment. If a yellow light appears, return it to the
charging case. Refer to the LED indicator on the
next page or on the charger case bottom label.
2. Place the OrthoPulse® device in your mouth, cen-
tering it between the front teeth.
3. Bite down gently to hold it in place. The device
will beep twice and the status light will turn blue
indicating that the treatment has started. A warm,
pleasant sensation can be felt during treatment.
The device has temperature management system,
if the temperature exceeds 48 °C the treatment
will be paused for cooling, the treatment will be
auto resumed after several seconds.
4. Once the treatment is complete, the device will beep three times
continuously and the blue status light will start pulsing.
5. Flip the device and repeat steps 2 through 4 to treat the other jaw.
6. Return the device to the plugged-in charging case to re-charge the
device after treatment.
TIP: Avoid loud background noise during treatment to ensure you hear
the aural indicators.
During the treatment, the device controller (box outside the mouth)
could reach a temperature above 41 °C, patient’s lip may contact the
device controller during the treatment. In case the heat exceeds the
patient’s comfort level the device can be removed from the mouth to
trigger the cooling mode by pausing the treatment.
Warning: Do not use the device under direct sunlight, or in an
environment with ambient temp above 30 °C.
There is no clinical benefit or any additional risk to the patient in the
case the device reaches 41 °C temperature. The device Firmware has
safety features that pause the treatment or cease the device opera-
tion in the case of overheating.