NESS L300 User Guide 3
For Your Health and Safety
Contraindications
Patients with a demand-type cardiac pacemaker should not use
the NESS L300.
Stimulation should not be applied over, or in proximity to,
cancerous lesions.
The NESS L300 should not be used over areas of regional
disorders, such as a fracture or dislocation, which would be
adversely affected by motion from the stimulation.
Warnings
The long-term effects of chronic electrical stimulation are
unknown.
The orthosis should not be applied over swollen, infected, or
inflamed areas or skin eruptions, e.g., phlebitis, thrombophlebitis,
varicose veins, etc.
Simultaneous connection of the L300 to the patient and to high-
frequency surgical equipment may result in burns at the site of
the stimulator electrodes and possible damage to the electronic
module of the orthosis.
Do not use the NESS L300 in close proximity (less than 3 feet) to
short wave or microwave therapy equipment as it may produce
instability in the orthosis electronic module output.
System configuration should only be performed by an authorized
clinician.