User's Manual

NESS L300 User Guide 33
6. Replacing and Installing
System Components
Replacing and registering components
When one or more of the NESS L300 electronic components are
replaced, a procedure must be performed to register
the new component
as part of the system. This ensures that there is radio communication
between the components of the system. These components are:
Stimulation unit
Foot sensor
Electronic control unit
1. Make sure the control unit is turned OFF.
2. Press the “Trigger mode button” and “-” button at the same time
and hold them for 3 seconds.
The segments of the stimulation intensity display will light up in 2
alternating circular patterns until the registration is complete (See
Figure ).
The letter “C” (short for connected) will appear if registration was
successful and a green light will appear on the control unit indicating
a successfully registered component
Note
If the component registered was the orthosis'
electronic module its status light will also turn green.