User Manual

112
NESS L300 Plus
Users Guide
Stimulation works in training mode but not in gait mode. When
I turn on gait mode I hear a beep, the RF Stim Units and Gait Sensor
indicators on the Control Unit alternately flash RED, and “E” flashes
in the digital display.
The Gait Sensor and RF Stim Unit are not communicating. The
Gait Sensor is probably asleep. Apply pressure to the Gait Sensor
pressure sensor. If this does not resolve the problem, the battery
may be depleted or the Gait Sensor may be faulty. If no wire issues
are apparent, replace the Gait Sensor battery and try again.
When I turn on the Control Unit, it beeps, the Control Unit and
RF Stim Unit indicators alternately flash RED, and “E” flashes in the
digital display. The RF Stim Unit indicators are not lit.
The RF Stim Unit battery is likely discharged, preventing the Control
Unit and RF Stim Unit from communicating. Turn off the Control
Unit, and charge the Control Unit and RF Stim Unit fully. Then,
disconnect the charger and turn on the Control Unit. The Control
Unit On/Off button and the status light on the RF
Stim Unit should flash GREEN. Communication should be restored.