I tried the registration procedure and saw a “C” immediately, but I
never saw the alternating arches in the digital display. The replacement
component is not working.
Clinician mode (for use by clinicians only) may have been started
instead of the registration process. Clinician mode is started by
pressing the minus and on/off buttons on the Control
Unit. Registration is started with the Control Unit off, and then by
pressing the minus and mode buttons on the Control
Unit. Turn off the Control Unit, and press the minus and
mode buttons to restart the registration process.
The Control Unit (or RF Stim Unit) does not light up when turned on.
The battery needs to be charged. Charge the battery. If the problem
persists, contact Bioness.
After I fully charged the Control Unit and RF Stim Unit, I disconnected
and then immediately reconnected the system charger set. The
charging icons displayed again. Are the components still fully
charged or do I need to repeat the charging process?
If you just charged your system and the fully charged icons
were displayed, your system is still fully charged. You do not
have to repeat the charging process.