User Manual

118
NESS L300 Plus
Users Guide
RF Stim Unit Specifications
Classification
Internally powered, continuous operation with type BF
applied parts
Operating
Voltage
3.7 V
Battery Type Proprietary rechargeable Li-Ion (Lithium Ion) 3.7 V, 700 mAh
Indications
Status (fault, battery, charging) and Stimulation LEDs
“Beeps” for audio alerts
Dimensions
Length: 74 mm (2.9 in.)
Width: 43 mm (1.7 in.)
Height: 15 mm (0.6 in.)
Weight 50 grams (1.6 oz.)
Environmental
Ranges
Transport and storage temperature: -20°C to +60°C (-4°F to
+140°F)
Operating conditions temperature: 5°C to 40°C (41°F to
104°F)
Charging temperature: 5°C to 40°C (41°F to 104°F)
Relative humidity: 25% to 85%
Atmospheric pressure: 900 hPa to 1060 hPa
Pulse Parameters
Pulse Balanced Biphasic
Waveform Symmetric or Asymmetric
Intensity 0–80 mA, 1-mA resolution (positive phase)
Max Voltage 120 V