User's Manual Part 6

129
Chapter 12 - Technical Specifications
Symmetric Asymmetric
Positive Pulse
Duration (µsec)
100 200 300 100 200 300
Negative Pulse
Duration (µsec)
100 200 300 400 800 1200
Inter-Phase
Interval (µsec)
50 0
Total Pulse
Duration (µsec)
250 450 650 500 1000 1500
Max Load 5000 ohm (Subject to max voltage limitation)
Pulse Repetition
Rate
20–45 Hz, 5-Hz resolution
Gait Parameters
Swing Control
Delay (%)
0–100% of phase* time, 5% resolution
Swing Control
End (%)
0–100% of phase* time, 5% resolution
Stance Control
Delay (%)
0–100% of phase* time, 5% resolution
Stance Control
End (%)
0–100% of phase* time, 5% resolution
Ramp Up 0–2 seconds, 0.1-second resolution
Ramp Down 0–2 seconds, 0.1-second resolution
Extend (%) 0–100% of stance time, 10% resolution
Max. Duration of
Stimulation
2–10 seconds, 1-second resolution
* Stimulation burst can start either on swing or stance phase.