User's Manual Part 6

137
Chapter 13 - Appendix - EMI Tables
Guidance and Manufacturers Declaration—
Electromagnetic Immunity for All Equipment and Systems
The NESS L300 Plus System is intended for use in the electromagnetic environment
specified below. The customer or the user of the NESS L300 Plus System should
assure that it is used in such an environment.
Immunity Test IEC 60601 Test
Level
Compliance Level Electromagnetic
Environment—
Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
6 kV contact
8 kV air
6 kV contact
8 kV air
Floors should be
wood, concrete, or
ceramic tile. If floors
are covered with
synthetic material,
the relative humidity
should be at least
30%.
Electrical fast
transient/burst
IEC 61000-4-4
2 kV for power
supply lines
1 kV for input/output
lines
2 kV for power
supply lines
Mains power quality
should be that of a
typical commercial
or hospital
environment.
Surge
IEC 61000-4-5
1 kV line to line
2 kV line to earth
1 kV line to line
(Class II without
any grounded
interconnections)
Mains power quality
should be that of a
typical commercial
or hospital
environment.