User's Manual Part 6

139
Chapter 13 - Appendix - EMI Tables
Guidance and Manufacturers Declaration—Electromagnetic Immunity
The NESS L300 Plus System is intended for use in the electromagnetic environment
specified below. The customer or the user of the NESS L300 Plus System should
assure that it is used in such an environment.
Immunity Test IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment—
Guidance
Portable and mobile RF
communications equipment
should be used no closer to
any part of the NESS L300 Plus
System, including cables, than
the recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to
80 MHz
3 Vrms
150 kHz to
80 MHz
Recommended separation
distance:
d = 1.2√P
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to
2.5 GHz
[E
1
] = 10 V/m
in 26 MHz to
1 GHz
[E
1
] = 3 V/m
in 1 GHz to
2.5 GHz
Recommended separation
distance:
d = 0.4√P, 80–800 MHz range
d = 0.7√P, 800–1000 MHz range
d = 2.3√P, 1000–2500 MHz range