Chapter 10 - Electronic Registration
To register a replacement/additional Intelli-Sense Gait Sensor:
Note: Read steps 1–4 before starting.
1. Turn off the L300 Plus System.
2. Simultaneously press and hold for three seconds the mode and
minus buttons on the L300 Plus Control Unit. The L300 Plus Control Unit
will beep when registration begins.
3.
The L300 Plus Control Unit digital display will show two ALTERNATING
GREEN arches while registration is in process.
4.
Within 15 seconds of initiating the registration procedure, repeatedly
press and release the pressure sensor.
5. When registration is complete, (“C” for complete) will appear in the
digital display, the Intelli-Sense Gait Sensor indicator will turn GREEN for
a few seconds, and the L300 Plus Control Unit will beep. If registration
fails, then repeat the procedure.
Note: If (“E” for error) appears in the digital display, an error has occurred.
Repeat the registration procedure. ( "E" also can mean that the registration
procedure was successful on a prior attempt and not noticed.)
6. After registration is complete, turn on the NESS L300 Plus System and
select gait mode. Press and release the Intelli-Sense Gait Sensor pressure
sensor. If the new Intelli-Sense Gait Sensor is registered, the mode
button will FLASH YELLOW rapidly for four seconds.