User's Manual

Guidance and Manufacturers Declaration Electromagnetic Emissions
The StimRouter system is intended for use in the electromagnetic environment specified below.
The customer or the user of the StimRouter system should assure that it is used in such an
environment.
very low, about 100 times lower than a commercially
88
Clinicians Guide
Emissions Test
Compliance
Electromagnetic Environment - Guidance
RF emissions
CISPR 11
Group 1
The StimRouter system uses RF energy for short-
range communications. Therefore, its RF emissions are
available cell phone. Though unlikely, portable and
mobile RF communications equipment, such as the
StimRouter patient programmer, EPT and clinician
programmer, could affect medical electrical equipment.
The RF-enabled components of the StimRouter system
are suitable for use in all establishments, including
domestic establishments and those directly connected
to the public low- voltage power supply network that
supplies buildings used for domestic purposes.
RF emissions
CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies