Guidance and Manufacturers Declaration Electromagnetic Immunity
The StimRouter system is intended for use in the electromagnetic environment specified below.
The customer or the user of the StimRouter system should assure that it is used in such an
environment.
Environment - Guidance
d = 1.2√P
800 Mhz
2.5 GHz
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2:
These guidelines
may not
apply
in all
situations. Electromagnetic propagation
is affected
by absorption and reflection from structures, objects and people.
90
Clinicians Guide
Immunity Test
IEC 60601 Test
Level
Compliance
Level
Electromagnetic
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80
MHz
3 V/m
80 Mhz to 2.5
GHz
3 Vrms
150 kHz to 80
MHz
10 V/m
26 MHz to 1 GHz
3 V/m
1 GHz to 2.5 GHz
Portable and mobile RF
communications equipment
should be used no closer to
any part of the StimRouter
system, including cables, than
the recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance (d)
d = 0.4√P 80 MHz to
d = 2.3√P 800 MHz to
Where P is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey,a should be less than
the compliance level in each
frequency range.
b
Interference
may occur in the vicinity of
equipment marked with
the following symbol: