User's Manual

Programming Parameters
Patients require tailored stimulation patterns to help control their pain. The StimRouter system
features eight programmable parameters and can store up to eight stimulation programs on the
clinician programmer, patient programmer and EPT. Timing parameters are specified in Table
5-4. Pulse parameters are specified in Table 5-5.
Table 5-4. Timing parameters.
30
Clinicians Guide
Parameter
Specification
Pulse
Balanced biphasic (pulse is hardware balanced no DC
component exists)
Waveform
Symmetric or Asymmetric
Intensity*
0-30 milliamperes peak, 1-milliampere resolution (positive
phase)
Maximum Voltage
100 volts
Maximum Output
7 milliamperes root mean square, 40 volts root mean
square
Maximum Charge Allowed
10 microcoulombs per phase
Electrode Current Density
Less than 1 milliampere root mean square per
centimeter2
Positive Phase Duration**
70, 100, 150, 200, 250, 300, 350, 400, 450, 500
microseconds
Negative Phase Duration
Symmetric: Identical to the positive phase duration.
Asymmetric: Four times the positive phase duration.
Parameter
Definition
Specification
On Time
Time that stimulation is
applied per cycle
1-60 seconds, 1-second
resolution
Off Time
Time that stimulation is
turned off per cycle
0-60 seconds, 1 second
resolution (0 seconds =
constant stimulation)
Ramp Up
Time to increase stimulation
from zero to the set intensity
0-10 seconds, but not more
than “On time”, 1- second
resolution
Total Time
Duration from the initiation
to the end of a stimulation
program
10 minutes- 12 hours