User's Manual

CAUTION: Make certain the adhesion site is free of obstructions (for example, bandages,
clothing, etc.) before adhering the StimRouter electrode. If the StimRouter electrode is placed
partially or wholly over a bandage or other obstruction, then skin irritation or tissue damage could
occur during stimulation.
7. Firmly adhere the StimRouter electrode to the skin, making sure that the electrode is in
full contact with the skin. If the StimRouter electrode is not firmly adhered to the skin
and moves, then stimulation may become uncomfortable or ineffective.
8. Make certain that the patient programmer is within 10 feet of the StimRouter electrode
(with EPT attached).
Confirming Set-Up
If the clinician programmer, patient programmer, EPT are connected correctly, the clinician
programmer information icon will be GREEN, confirming online mode.
To confirm the clinician programmer is in online mode:
1.
2.
Check that the information icon on the clinician programmer is GREEN.
If the information icon is not GREEN, make certain that the StimRouter electrode with
the EPT attached is within 10 feet of the patient programmer.
If the
information
icon is still not
GREEN, check
all
connections
and charge the patient
programmer and the EPT.
Test the EPT using the tester.
See the Troubleshooting section of this guide.
3.
4.
5.
6.
Refer to the StimRouter Users Guide for additional operational instructions for the StimRouter
patient programmer, EPT and StimRouter electrode, including instructions on how to:
Turn on the patient programmer.
Charge the patient programmer.
Adjust the volume of audio alerts on the patient programmer.
Select a stimulation program using the patient programmer.
Turn stimulation on using the patient programmer.
Chapter 6 - Set-Up Instructions