User's Manual

Testing the EPT
The tester is used to test the functionality of an EPT. Before connecting the EPT to the tester,
make
sure the
EPT
is
charged. Using
the
patient programmer
or
clinician programmer,
stimulation
can be turned on.
Using the tester:
1.
Connect the Tester snaps to the EPT plug holes. See Figure 6-14.
Figure 6-14. Connecting the tester to the EPT.
2.
3.
Press the patient programmer on/off button to turn on the system.
Press the mode button briefly to start stimulation, you should hear a buzzing coming from
the tester when stimulation is on and no buzzing when stimulation is off.
En
Chapter 6 - Set-Up Instructions
43