User's Manual

Note: The stimulation settings have a constant ramp-up of 1 second, during which the patient
may not feel any stimulation. Account for ramp-up time during the adjustment process.
4. Press Test to test the stimulation settings; stimulation will turn on.
Note:
If the combination of parameters exceeds the maximum charge level allowed (10 microcoulombs
per phase), then a warning will appear with directions to reduce the total charge. “Test” will be
disabled. See Figure 9-2.
Figure 9-2. Maximum charge level exceeded.
5.
Slowly increase the intensity to a level that is tolerable for the patient and achieves the
desired therapy.
Press
Stop &
Save
to stop
stimulation
and save the
stimulation settings
to the
clinician programmer, patient programmer and EPT. See Figure 9-3.
6.
Figure 9-3. Press “Stop & Save” to stop testing and save the current stimulation settings.
66
Clinicians Guide