User's Manual

3. As
needed,
use
additional saturated disinfectant wipes
or
cloths
to keep the component
surface wet for 10 minutes.
Note: Be sure to follow the Bioness instructions for the specified contact time to ensure
an effective bacteria kill.
Clinician Kit Carrying Case
The StimRouter Clinician Kit carrying case may be cleaned and low-level disinfected using
CaviCide® (Metrex, Orange, CA) or 70% isopropyl alcohol (IPA) per the instructions below:
CaviCide:
1.
2.
Spray the entire surface of the Clinician Kit carrying case with CaviCide.
Use a clean towel to remove any surface contaminants. Soil, etc., will impede the
disinfectants effectiveness, if not removed.
Spray the entire surface of the Clinician Kit carrying case again with CaviCide.
Keep spraying the entire carrying case surface as needed to keep it wet for 10 minutes.
3.
4.
70% IPA:
1.
Wipe the entire surface of the Clinician Kit carrying case with a wipe or cloth saturated
with 70% IPA.
Use a new wipe or cloth saturated with 70% IPA to remove any surface contaminants.
Soil, etc., will impede the disinfectants effectiveness, if not removed.
Wipe the entire surface of the Clinician Kit carrying case again with a new wipe or cloth
saturated with 70% IPA.
Use additional new wipes or cloths as needed to keep the entire surface of the carrying
case wet with 70% IPA for 10 minutes.
2.
3.
4.
Note: Be sure to follow the Bioness instructions for the specified contact time to ensure an
effective bacteria kill.
Note: Do not use other cleaning/disinfecting agents such as a diluted bleach mixture, Lysol or
Clorox wipes. Bioness has not tested these products effectiveness
on
the StimRouter
components.
Metrex products are sold through authorized dealers worldwide; 70% IPA is widely available. If
you need assistance, please contact your local distributor.