User's Manual

Note: The patient Usage Log is stored on the EPT and is
automatically
retrieved when the patient
programmer and EPT are connected.
11. Proceed with the programming session. Programming changes will be automatically
saved to the EPT.
12. Before the session ends, remind the patient to re-register his or her EPT and patient
programmer at home. Review with the patient the following instructions for RF registration.
Registering the Patient Programmer/EPT
The patient programmer and EPT must be electronically registered to each other to operate.
To register the patient programmer/EPT:
1.
2.
If necessary, charge the patient programmer.
Ensure that the patient programmer is turned off (the on/off button should not be
FLASHING GREEN). See Table 11-1.
Table 11-1. Patient programmer on/off indicators.
3.
4.
Make sure the EPT is charged (if not, connect the EPT to a charger).
Place the patient programmer and the EPT close together on a table. Make sure the
patient
programmer
and the EPT are no more than a few inches apart and not
touching.
Make certain that any other EPT is at least 10 feet away from the components to be
registered.
Simultaneously press and hold for three seconds the mode and minus buttons on the
patient programmer to start the registration process. See Figure 11-2. The patient
programmer will sound an audio alert, indicating the registration process has begun.
The patient programmer digital display should show two ALTERNATING GREEN
ARCHES, indicating registration is in progress.
5.
6.
7.
Chapter 11 - Troubleshooting
75
Patient Programmer OFF
Patient Programmer ON
Flashing Green