User's Manual

On/Off Button
Mode Button
Volume Buttons
Plus/Minus Buttons
Unregistered
In Process
Complete
Error
Figure 11-2. StimRouter Patient Programmer operating buttons and RF registration status indicators.
8. When the digital display shows the letter “C and the RF icon turns GREEN for a few
seconds, the registration procedure is complete.
Note: If the digital display shows the letter “E” and the RF icon turns RED for a few seconds,
then an error has occurred and the procedure must be repeated. If the problem persists,
reconnect the charger to the EPT. If the error indication continues to appear, see Table 11-7.
Patient Forgets EPT
If a patient forgets to bring his or her EPT to a follow-up session, reschedule the programming
session. Do not substitute a clinic EPT for the patients EPT; only the patient programmer can
be substituted. The information stored on an EPT will always overwrite the information stored
on a patient programmer. If a clinic EPT is used, then the patients EPT will overwrite any new
information stored on the patient programmer.
76
Clinicians Guide