User's Manual

Patient Forgets StimRouter Electrode
If a patient forgets to bring his or her StimRouter electrode to a follow-up session, use a clinic
StimRouter electrode or reschedule the programming session
Troubleshooting Wireless Technology
Table 11-2. Troubleshooting, clinician programmer.
78
Clinicians Guide
Clinician Programmer
Problems/Solutions
FLASHING RED Information
Icon
A User-Correctable Error such as Faulty
Electrode Contact, Radio Communication
Failure
Press the information icon to view the error
message and list of solutions.
FLASHING YELLOW
Information Icon
Low Battery Detected: EPT or Patient
Programmer
• Press the information icon to view the error
message and list of solutions.
• Charge the patient programmer.
• Charge the EPT.
RED Information Icon
Software or Hardware Error
• Press the information icon to view the error
message and list of solutions.
Unexpected Error in the StimRouter
Application
Contact your local distributor.
Clinician Programmer Will Not Turn On
Press the clinician programmer reset button.
Charge the clinician programmer and verify
that the amber LED is ON. If the light does not
turn on, refer to the HP instructions for use.
Contact your local distributor.