Table 11-5. Troubleshooting, patient programmer, EPT icon.
80
Clinicians Guide
EPT Icon
Problems/Solutions
EPT Battery Charge Level is Low
Charge the EPT.
EPT Icon FLASHES
YELLOW
EPT Icon GLOWS RED and
“E” appears in the Digital
Display
EPT Malfunction
Contact your local distributor.
EPT Icon FLASHES RED
and “E” FLASHES in the
Digital Display
EPT Temperature Error
The EPT is either too hot or too cold and will
range is restored. Restore the EPT to its
working temperature range.
EPT Icon FLASHES
RED and Intensity Level
FLASHES in the Digital
Display
Faulty Electrode Contact
Turn off the patient programmer.
skin.
removed from the StimRouter electrode.
Check to see that the EPT is correctly
attached to the StimRouter electrode. If
necessary, reattach the EPT into the
StimRouter electrode.
Make sure that nothing is on the gel electrodes
or on the skin that would interfere with
electrode contact with the skin.
Re-adhere the StimRouter electrode to the
skin.
Replace the StimRouter electrode.
Clean the skin.