Table 11-6. Troubleshooting, patient programmer, patient programmer icon.
the patient programmer.
patient programmer but not touching.
Chapter 11 - Troubleshooting
81
RF Icon
Problems/Solutions
RF Icon GLOWS RED and
“E” Appears in the Digital
Display Immediately After the
Registration Attempt
Registration Failure; Unknown Reason
• Retry the process.
RF Icon GLOWS RED, and “E” and then
“2” Appear in the Digital Display
Immediately After the Registration Attempt
Registration Failure; More than One EPT Found
• Make sure only one EPT is within 10 feet of
• Retry the process.
RF Icon GLOWS RED, and “E” and then
“0” Appear in the Digital Display
Immediately After the Registration Attempt
Registration Failure; No EPT Found
• Make sure the EPT is within inches of the
• Connect the EPT to the charger.
•
• Retry the process.
Patient Programmer Icon
Problems/Solutions
While Charging, Patient
Programmer Icon GLOWS
RED and “E” Appears in the
Digital Display
Charging Error
• Reconnect the charger cable to the patient
• Replace the battery in the patient programmer.
Patient Programmer Icon
GLOWS RED and “E”
Appears in the Digital Display
Patient Programmer Malfunction
• Turn the patient programmer off and then back
• Check to see if one of the buttons is stuck
and, if so, try to release it.
• Contact your local distributor.
Patient Programmer Icon
FLASHES YELLOW
Patient Programmer Battery Charge Level is
• Charge the patient programmer.