Table 11-7. Troubleshooting, patient programmer, RF icon.
Table 11-8. Troubleshooting, EPT and stimulation.
82
Clinicians Guide
EPT and Stimulation
Solutions
Stimulation Not As Effective As Usual
Check the orientation of the StimRouter
electrode, it should be over the receiver end of
the lead.
Make sure the StimRouter electrode is
securely adhered to the skin.
Visually inspect the StimRouter electrode to
make sure the gel is aligned with the contour
of the electrode.
Review the skin care instructions.
Clean the skin with a damp cloth.
Change the StimRouter electrode, if the skin
is dry.
Trim hair from the StimRouter electrode site.
Test the EPT using the tester.
Undesirable Motor Response
Decrease the stimulation intensity level.
Check the placement of the StimRouter
electrode.
EPT Charging Light Does Not Turn On
Check the connection.
Check the Y cable (disconnect it and connect
charger directly).
Check that the charger light turns on.
Contact your local distributor.
RF Icon
Problems/Solutions
RF Icon FLASHES RED
Radio Communication Failure or EPT Battery
Make sure the patient programmer and EPT
are within 10 feet of each other.
If the components are within range, then turn
the patient programmer off and back on.
Connect the EPT to the charger.• Re-register
the components.
Contact your local distributor