User's Manual
Failure
Table 11-7. Troubleshooting, patient programmer, RF icon.
Table 11-8. Troubleshooting, EPT and stimulation.
82
Clinician’s Guide
EPT and Stimulation
Solutions
Stimulation Not As Effective As Usual
• Check the orientation of the StimRouter
electrode, it should be over the receiver end of
the lead.
• Make sure the StimRouter electrode is
securely adhered to the skin.
• Visually inspect the StimRouter electrode to
make sure the gel is aligned with the contour
of the electrode.
• Review the skin care instructions.
• Clean the skin with a damp cloth.
• Change the StimRouter electrode, if the skin
is dry.
• Trim hair from the StimRouter electrode site.
• Test the EPT using the tester.
Undesirable Motor Response
• Decrease the stimulation intensity level.
• Check the placement of the StimRouter
electrode.
EPT Charging Light Does Not Turn On
• Check the connection.
• Check the Y cable (disconnect it and connect
charger directly).
• Check that the charger light turns on.
• Contact your local distributor.
RF Icon
Problems/Solutions
RF Icon FLASHES RED
Radio Communication Failure or EPT Battery
• Make sure the patient programmer and EPT
are within 10 feet of each other.
• If the components are within range, then turn
the patient programmer off and back on.
• Connect the EPT to the charger.• Re-register
the components.
• Contact your local distributor