User's Manual
12
Technical
Specifications
Wireless (RF) Communications Specifications
Chapter 12 - Technical Specifications
83
Capabilities
• Communication between the patient programmer and EPT
• Communication between the Clinician Programmer and Bluetooth-
enabled printer
Functions
RF Communication is used between the EPT and patient programmer,
enabling the Clinician Programmer to perform the following functions:
• Define stimulation parameters and programs (e.g., pulse width, amplitude)
• Control stimulation (e.g., start/stop stimulation, change programs,
adjust stimulation amplitude)
• Display system status (e.g., battery charge level, error codes)
• Retrieve patient stimulation session data
Bluetooth communication is used between the clinician programmer and
the HP printer to print stimulation parameters and use history.
Modes
• Online Clinician Programmer mode (Clinician Programmer connected to
patient programmer and EPT)
Characteristics
EPT and patient programmer
• Frequency band: 2.400 up to 2.4835 GHz
• RF Frequency channels: 29 channels
• Channel spacing: 580.810 kHz
• Antenna type: Integrated chip antenna
• Transceiver duplexing scheme: TDD
• Frequency synthesizer settling time: < 1 msec
• Modulation frequency: 0.5 Mbps
• Modulation type: Minimum Shift Keying (MSK)
• Transmit power (EIRP): EPT, -16 dBm; patient programmer, 4.1 dBm
• Modulation baud rate: 500 kHz
• Modulation bandwidth: 590 kHz
Clinician Programmer
• Bluetooth 2.0