User's Manual

12
Technical
Specifications
Wireless (RF) Communications Specifications
Chapter 12 - Technical Specifications
83
Capabilities
Communication between the patient programmer and EPT
Communication between the Clinician Programmer and Bluetooth-
enabled printer
Functions
RF Communication is used between the EPT and patient programmer,
enabling the Clinician Programmer to perform the following functions:
Define stimulation parameters and programs (e.g., pulse width, amplitude)
Control stimulation (e.g., start/stop stimulation, change programs,
adjust stimulation amplitude)
Display system status (e.g., battery charge level, error codes)
Retrieve patient stimulation session data
Bluetooth communication is used between the clinician programmer and
the HP printer to print stimulation parameters and use history.
Modes
Online Clinician Programmer mode (Clinician Programmer connected to
patient programmer and EPT)
Characteristics
EPT and patient programmer
Frequency band: 2.400 up to 2.4835 GHz
RF Frequency channels: 29 channels
Channel spacing: 580.810 kHz
Antenna type: Integrated chip antenna
Transceiver duplexing scheme: TDD
Frequency synthesizer settling time: < 1 msec
Modulation frequency: 0.5 Mbps
Modulation type: Minimum Shift Keying (MSK)
Transmit power (EIRP): EPT, -16 dBm; patient programmer, 4.1 dBm
Modulation baud rate: 500 kHz
Modulation bandwidth: 590 kHz
Clinician Programmer
Bluetooth 2.0