User Manual

Table Of Contents
Operating Instructions 8-35
The saved level will be the default level whenever the stimulation
program is started or until a new default level is saved.
Turning Stimulation Off
Stimulation Off Stimulation On
To turn stimulation off:
1 Press the mode button briefly. The remote control will return to
standby mode and the mode button will turn off. Stimulation will
stop.
Remote Control Off Remote Control On
2
Or, turn the remote control off. The remote control can be turned off
at all times. When the remote control is turned off, stimulation will
stop.
Note:
The remote control is designed to mitigate unintentional stimulation
or stimulation at an intensity level that is too high. If stimulation is
inadvertently set too high or turned on unintentionally, and cannot be
corrected using the remote control, remove the user patch. If the remote
control malfunctions, or an unexpected response or stimulation level is
experienced, take the user patch off and contact Bioness Customer
Support.