A cancerous lesion near the target stimulation area or near to where
the user patch will adhere
A fracture or dislocation near the target stimulation area or near to
where the user patch will adhere
The StimRouter should not be used if you may require:
Diathermy treatment
Magnetic resonance imaging
Warnings
Diathermy
Shortwave, microwave and/or therapeutic ultrasound diathermy should
not be used on patients who have a StimRouter Implantable
Neurostimulator. The energy generated by diathermy can be transferred
through the StimRouter, causing tissue damage at the lead site and
resulting in severe injury or death.
Diathermy may also damage the neurostimulation components, resulting
in loss of therapy and requiring additional surgery for lead replacement.
Injury or damage can occur during diathermy treatment whether the
neurostimulation system is turned on or off. All patients are advised to
inform their health-care professional that they should not be exposed to
diathermy.
Magnetic Resonance Imaging (MRI)
Patients implanted with the StimRouter Implantable Neurostimulator
should not be subjected to MRI. MRI exposure may dislodge the
implanted lead, heat the lead, damage the component electronics and/or
cause voltage induction through the lead.
Implanted Stimulation Devices
Neurostimulators, such as the StimRouter, may interfere with the
operation of implanted sensing stimulators such as pacemakers or