8-34 StimRouter User Manual
this level will be most appropriate for use. Stimulation intensity can be
adjusted, however, using the plus/minus buttons.
CAUTION:
Be careful not to set the stimulation intensity too high. Pain can result.
To temporarily adjust stimulation intensity:
1 Either from standby mode or while stimulating in stimulation mode,
press the plus or minus button once for each level of change. The
remote control will beep and the new level will show on the digital
screen.
2 Allow time for the system to demonstrate a change in setting before
making another change.
CAUTION:
If the intensity level is changed while stimulation has cycled off in
stimulation mode, or while the system is in standby mode, the new
intensity level will not activate until the system cycles back on. Be certain
to test each level before advancing stimulation to a different level,
especially to a higher level.
Saving a New Default Stimulation Intensity Level
You can save a new default stimulation intenstity level for each program
stored on the remote control.
To save a new default stimulation intensity level:
1 Select a stimulation program (“Selecting a Stimulation Program” on
page 8-32).
2 Return to standby mode.
3 Adjust the stimulation intensity level to the desired level
(“Adjusting Stimulation Intensity” on page 8-33).
4 From standby mode, press and hold the minus button for at least
three seconds. The remote control will beep and the new default
intensity level will flash on the digital display for a few seconds
while saving.