Testing the Position of the L300 FS Cuff
To test the position of the L300 FS Cuff:
1. Place the L300 Plus Control Unit in standby mode.
2.
Press the L300 RF Stim Unit selection button. The arrow next to the L300
RF Stim Unit indicator will light GREEN to confirm the selection.
3. Press and hold the stimulation test button. The L300 RF Stim Unit will
stimulate until the stimulation test button is released.
If the L300 FS Cuff is not positioned correctly:
1. Turn off the L300 Plus System.
2. Unhook the L300 FS Cuff strap handle from the cradle.
3. Slowly lift the L300 FS Cuff away from the skin.
4. If appropriate, gently peel the L300 hydrogel electrodes from the skin.
Then make sure the L300 hydrogel electrodes are adhered correctly to
the bases.
5. Repeat the steps for putting on the L300 FS Cuff.
6. Retest the position of the L300 FS Cuff.
CAUTION: Remove the L300 FS Cuff from the skin every 3 to
4 hours for 15 minutes to allow the skin to breathe.