User's Manual

41
40
About Quality Control Testing
Performing a Quality Control Test
CAUTION
- Each time you open a new bottle of control solution, write the expiration date on the label. The
is good for 3 months after opening the bottle, or until the expiration date printed on
the label, whichever comes first.
Rightest
Control Solution GC550
- Wipe the bottle cap with a clean tissue before tightly closing the bottle of control solution.
- Close the bottle of control solution tightly immediately after using.
- Check the expiration date before use. (Refer to the Package Insert).
- Keep control solution bottles out of reach of children.
Rightest Control Solution GC550
Example
Exp. Date 2015-06
1. Take one Rightest Blood Glucose Test Strip from the vial and close the vial cap
immediately.
2. Insert the Blood Glucose Test Strip horizontally (not vertically) with the
colored view window facing up into the test strip port.
3. While the blood drop icon is flashing on the display window, press and hold the Main button
for at least 3 seconds until the " " symbol appears.
4. You will see the blinking " " icon and " " icon on the screen, prompting you to apply
.
5. Shake the bottle of well before opening the cap. Place the
cap on a flat surface.
6. Place a drop of control solution onto the top of the cap.
7. Gently touch the sample port of the strip with the control solution from the top of the cap.
8. The screen will display the count time starting from 5 (you will hear a beep if the volume is
turned on).
GM232B
GM232B Rightest
Rightest Control Solution GC550
Rightest Control Solution GC550
8
3
OPEN
5
6
1
2
7