User's Manual

9. Tightly replace the cap on the bottle.
10. The control solution result will appear. Compare your Quality Control Test result to the Control Solution
Range printed on the Rightest Blood Glucose Test Strip vial label.
Rightest Control Solution GC550
GM232B
If Rightest Control Solution GC550 results are out of range, your Rightest Blood Glucose Monitoring System GM232B may
not be working properly. Repeat the Quality Control Test. If your control solution results are still out of range, do not use
Rightest Blood Glucose Meter to test your blood glucose. Please contact the GE Diabetes Customer Support
Center .
GM232B
toll free at 888-481-8485
43
Understanding Control Test Results
42
Performing a Quality Control Test
CLOSE
9
CAUTION
- Your Control Solution Test results will not be included in the average calculations, however, they can still can be recalled
and viewed. The Control Solution Test result will be shown with the " " icon on the screen.
- The Control Solution Test should be conducted between 59-104°F (15-40°C).
- Before " " and " " appear, do not touch the control solution to the sample port on the strip. The Rightest Blood
Glucose Meter is performing an internal check. Touching the control solution to the sample port before
prompted will result in an error message: " " and " " .
- Do not drip the control solution on to the sample port of the test strip directly. The reagent on the strip might leak into
the bottle of control solution and may cause the degeneration of the control solution. This could
contaminate the meter via the test strip port.
- Keep the test strip port clean and dry. Clean immediately if the test strip port is stained or is overly
exposed to moisture.
- Do not touch the tip of the control solution bottle. If the tip is touched, clean with water.
GM232B
and be accompanied by beeps (if volume is turned on)
Your control solution test results should fall within the control solution range printed on the test strip vial label. If the
results are within the range, the Rightest Blood Glucose Monitoring System GM232B is working correctly.
Possible reasons your Control Solution results are out of the range :
- Your is expired or was first opened more than 3 months ago.
- Your Rightest Blood Glucose Test Strip has expired.
- You left the cap of the Blood Glucose Test Strip vial or the control solution off for a period of time.
- You did not perform the test procedure correctly.
- The Blood Glucose Meter or Blood Glucose Test Strip have malfunctioned.
Rightest Control Solution GC550
Rightest Rightest
GM232B
GM232B GM232B
Example of control solution range
printed on your test strip vial label.
Control Solution Range
88~120 mg/dL
250~340 mg/dL
Level 2
Level 4